Enlargement Tables
           graphic             .graphic
ZOOMING
Since this is an indirect vision system, zoom factors may vary to suit the  distance of the endoscope from the object being filmed and the monitor used. 
 
 
 
 
Distance
Enlargement 
mm
17" Screen 
19" Screen 
25
18
20
30
15
17
35
13
14
40
11
13
45
10
11
50
9
10
55
8
9
60
8
8
65
7
8
70
6
7
.
Resolution:  0.065mm
Regardless of the type of monitor used, the  INSIDE system has a fixed resolution, i.e. the  minimum size clearly perceived at the distance of  25 mm
.
Working Distance: 
25 - 60 mm
By 'working distance' we mean the distance  between the object and the endoscope, offering  the best image quality.
If you focus on objects at a distance of more than  60mm, you get a 'fog' effect. This is normal as  you are working beyond the nominal working  distance.  
.
Automatic Brightness  Correction
The endoscope electronically controls changes in  outdoor light levels in order to guarantee  optimised images. 
INSIDE produces the light needed autonomously.